TabTabletop autoclaves are the backbone of sterilization for small medical offices, dermatology practices, dental clinics, and medspas. Compact yet powerful, they ensure instruments are sterilized safely, protecting both patients and providers.
But like any equipment, these autoclaves develop recurring issues — usually from daily wear, skipped maintenance, or environmental factors like poor water quality. The good news is that many of these problems can be diagnosed quickly, and in some cases, fixed without waiting on a service call.
Below we’ll walk through the most common autoclave problems, what causes them, and what to do next. Along the way, we’ll include real-world scenarios to show how these issues play out in the field.
1. Autoclave Won’t Power On
Imagine this: a dermatology clinic is getting ready for back-to-back procedures, but the autoclave won’t even light up. The front desk is panicking, because without sterilized instruments, the day’s schedule is on the line.
Common Causes:
- Loose power cable or faulty outlet
- Tripped circuit breaker or blown fuse
- Faulty door switch (some autoclaves won’t power unless the door is fully closed)
Fix:
- Confirm the outlet works by testing another device.
- Inspect the fuse or breaker and replace/reset as needed.
- Check that the door is latched tightly — on many tabletop units, an incomplete latch interrupts the power circuit.
- If it still won’t start, the control board or wiring likely needs professional repair.
2. Chamber Won’t Pressurize
One of the most common calls we get is: “The cycle runs, but the autoclave never builds pressure.” In one case, a medspa reported hearing faint hissing during every cycle — sure enough, the door gasket had cracked, and steam was escaping instead of building pressure.
Causes:
- Worn or cracked door gasket
- Loose or unevenly latched door
- Scale buildup around valves from using tap water instead of distilled
- Leaky fittings or solenoid valves
Fix:
- Replace door gaskets annually, or sooner if cracks or flattening appear.
- Confirm the door closes evenly — sometimes hinges loosen over time.
- Clean valves and fittings.
- Always use distilled water only to prevent scale buildup.
3. Sterilization Cycle Won’t Complete
A dental practice once reported their autoclave was aborting halfway through every cycle. After inspection, we found the drain filter was clogged with cotton fibers from gauze packs. Airflow was restricted, steam couldn’t circulate, and cycles failed.
Causes:
- Overloaded chamber (steam can’t circulate)
- Drain filter clogged
- Faulty temperature or pressure sensors
- Heating element wearing out
Fix:
- Load instruments loosely, allowing space for steam.
- Clean the drain filter weekly — it’s small but critical.
- If problems persist, sensors or heaters may require replacement.
4. Wet Packs After a Cycle
Few things frustrate staff more than opening the door to find wet, dripping packs. Not only is it unsafe (moisture can harbor microbes), but it means re-running the cycle — doubling sterilization time.
Causes:
- Chamber overloaded
- Improper wrapping material
- Drain line partially clogged
- Door gasket leaking steam at the wrong time
Fix:
- Reduce load size and leave room between packs.
- Only use wraps designed for autoclaves.
- Inspect drain lines and clean if blocked.
- Replace or lubricate the door gasket if leaking.
5. Strange Noises or Leaks
Another medspa complained about “knocking” sounds during cycles. After inspection, we found scale buildup inside the steam lines — mineral deposits from using tap water. The fix was a thorough descaling and a reminder to always use distilled water.
Causes:
- Solenoid valve beginning to fail
- Loose or worn chamber gasket
- Mineral buildup in chamber or tubing
Fix:
- Replace worn seals and valves.
- Perform regular descaling treatments (especially in hard-water areas).
- Call for service if noises persist — they can indicate valve timing issues.
6. Error Codes or “Service Required”
Modern tabletop autoclaves have error codes for safety. These can be intimidating, but often they point directly to the issue.
For example, an error might indicate a temperature sensor failure — meaning the unit isn’t sure how hot it actually is. In one dermatology office, a persistent error code turned out to be nothing more than a faulty door switch, replaced in under 30 minutes.
Common Codes Indicate:
- Sensor failure (temp/pressure)
- Electrical heating fault
- Board communication error
- Door/lid not sealed properly
Fix:
- Try a full power cycle.
- Refer to the user manual for code meaning.
- Some fixes are user-friendly (gaskets, switches), but persistent codes often need a technician.
Preventive Care to Avoid Most Problems
In almost every scenario above, the problem could have been prevented with basic care.
- Use distilled water only — tap water = scale buildup.
- Clean the drain filter weekly.
- Check gaskets monthly and replace annually.
- Avoid overloading to allow proper steam circulation.
- Schedule yearly preventive maintenance to catch small issues before they shut you down.
Final Thoughts
Tabletop autoclaves are dependable workhorses, but they’re not invincible. From leaking gaskets to wet packs, most problems follow predictable patterns and can be solved with simple checks.
By knowing what to look for — and addressing issues early — you can keep your autoclave running smoothly, minimize downtime, and protect patient safety.
And when the fix goes beyond basic troubleshooting? That’s where professional technicians step in. Having a trusted service partner means peace of mind that your autoclave is always ready when you need it most.
Looking for a Replacement?
Sometimes, repairing an older tabletop autoclave just isn’t cost-effective. If the heating element, control board, or chamber itself is failing, the repair costs may approach the price of a new unit.
That’s where Amtech Medical’s refurbished autoclaves come in.
We stock a range of used, professionally reconditioned autoclaves that are:
- Fully serviced and tested by our technicians
- Backed by warranty for peace of mind
- More affordable than new units — without sacrificing reliability
- Available in multiple models and sizes to fit the needs of dermatology, dental, histology, and medspa practices
Choosing a refurbished autoclave from Amtech Medical gives you the confidence of a fully inspected and serviced machine at a fraction of the cost of new. It’s the perfect option for practices that need dependable sterilization without stretching their budget.
👉 Explore our current inventory of used autoclaves on our website or reach out to our team for availability.
Professional Autoclave Service & Support
While some issues with tabletop autoclaves can be fixed in-house, others require a trained technician. Electrical failures, sensor replacements, valve timing issues, and control board repairs should always be handled by professionals to avoid bigger (and more expensive) problems down the line.
At Amtech Medical, we provide:
- Nationwide Service Coverage – Our technicians travel to support practices across the U.S.
- Preventive Maintenance Programs – Keep your autoclave reliable with scheduled inspections and cleanings.
- Emergency Repairs – Fast turnaround when your sterilizer goes down unexpectedly.
- Multi-Device Support – We service autoclaves alongside your other dermatology, lab, and general medical equipment.